Tel: +36-20-344-1787, E-mail: rapaskft@rapas.hu
5 and 10 kV insulation tester, ES3035+, ES3035E+
Range 2000 GΩ, resolution 0,01 MΩ. DC 0~1000 V, AC 0~750 V voltage measurement, DAR and PI calculation, USB interface
ES3035+ (5 kV) / ES3035E+ (10kV) insulationtester features
- Test Voltage (ES3035+): 100, 250, 500, 1000, 2500, 5000 V
- Test Voltage (ES3035E+): 250, 500, 1000, 2500, 5000 V, 10 kV
- Insulation Resistance Range: 0.01MΩ -2000GΩ
- Automatic discharge of objects after test
- Adjustable alarm limit
- Absorption ratio and PI measurement
- Short circuit current: ≥5 mA
- DC voltage range: 0-1000 V
- AC voltage range: 0-750 V
- Memory for: 500 groups of measured values
- USB interface + software
ES3035+ (ES3035E+) digital insulation resistance tester has a large backlit LCD display, data storage, data recall, limit value alarm, automatic discharge of the measured object and automatic power-off functions. In addition to the insulation test, the instrument calculates the absorption ratio (DAR)and the polarization index (PI).
The instrument with its high output power and high interference suppression operates in the measuring voltage range of 100V-5kV (ES3035+), or 250V-10kV (ES3035E+), while the insulation resistance measurement range for both types is from 0.01MΩ to 2000 GΩ, with the best resolution of 0.01MΩ. The measurement range of DC voltage is 0–1000 V and that of AC voltage is 0–750 V.
The memory of the ES3035+ (ES3030E+) digital insulation resistance tester can store 500 measured data groups downloadable to a PC by the USB interface and evaluated with the software supplied with the device.
The device has an impact, dust and moisture resistant case and can be used safely even in harsh working environments. Suitable for measuring the insulation resistance of various insulation materials, transformers, motors, cables and electrical equipment
ES3035+ (5 kV) / ES3035E+ (10kV)-os technical specification
- Insulation resistance measurement, ES3035+
- Test voltage: 100, 250, 500, 1000, 2500, 5000V ±10%
- Insulation resistance range: 0.01MΩ... 2000GΩ
- Resolution: 0.01 MΩ
- Accuracy: ±3% SD ±5 digits
- Insulation resistance measurement, ES3035E+
- Test voltage: 250, 500, 1000, 2500, 5000, 10,000V ±10%
- Insulation resistance range: 0.01MΩ... 2000GΩ
- Resolution: 0.01 MΩ
- Accuracy: ±3% SD ±5 digits
- Output short-circuit current: ≥5 mA (ES3035+: 5kV, ES3035E+: 10kV)
- DC voltage measurement
- Measuring range: 0.. 1000V
- Resolution: 0.1 V
- AC voltage measurement
- Measuring range: 0... 750V
- Resolution: 0.1V
- Measurement / display of absorption ratio and polarization index
- Alarm function: Alarm if the measured value is less than the alarm set limit value
- Power
- Power source:: 12.6 V rechargeable battery
- Power requirement: max. 30 mA (backlight off), : max. 42 mA max. (backlight on)
- Battery depletion indication: empty battery symbol on the display
- Automatic shutdown: after 15 minutes of idle
- Display
- Display: adjustable backlit 4-digit LCD, 108×65 mm
- Over-load indication: OL message on the display
- USB interface + software for downloading and evaluating measurement results to a computer
- Data storage
- Save 500 measured value groups in the memory
- Data can be called back to the display
- Dimensions / weight (L/W/D: 277.2x227.5x153 mm / 2720 g (with battery)
- Temperature ranges, humidity
- Use: -10... +50°C, ≤85% RH
- Storage: -15... 55°C, ≤90% RH
- High voltage test: AC 3kV/50Hz 1 minute
- Applicable standards: IEC61010-1, IEC1010-2-31, IEC61557-1.5, IEC60529(IP54) pollution degree 2, CAT III 300 V
Standard accessories
- 1 pc. high voltage probe
- 2 pcs. high voltage test lead (black, green)
- Monitoring software on CD
- 1 pc. USB cable
- 1 pc. battery charger
- User manual
- Letter of guarantee
- Carry bag
- Calibration protocol